Applications
and Features
Specifications
Principles
of Mapping
Sample
Mapping Report
Additional
Information Magnet
Management
|
MAGNETIC
FIELD MAPPING - NMR MAGNETS - FMU-1200
SPECIFICATIONS
- ELECTRONICS
|
NMR
spectrometer from 5 to 1280 MHz (0.1 to 30 Tesla) |
|
Phase-continuous
sweeps with resolution 1 Hz |
|
Computer
controlled frequency sweep rate up to 1MHz/s |
|
Sweep
range of 10 MHz |
|
Direct
digital synthesis (DDS) freque
ncy source |
|
Storage
and display capability for raw data, gradient maps, gradient
values, and gradient correction currents |
|
Digital
quadrature detection |
|
Over-sampling
ADC |
SPECIFICATIONS
- MAGNETICS
|
Complete
mapping and analysis of magnetic field |
|
Single
pass acquisition using helical mapping path |
|
Automatic
resonance detection |
|
Mapping
and characterization of correction gradients |
|
Analysis
capability for axial gradients z0 to z10 |
|
Analysis
capability for radial gradients x/y to c8/s8 |
|
Closed-loop
shim coil calibration (in communication with a programmable shim
system) |
|
Closed-loop
field correction (in communication with a programmable shim
system) |
|
Magnet
drift compensated measurements |
|
Cycle
time of less than 8 minutes |
SPECIFICATIONS
- MECHANICAL
|
Mounting
Hole Pattern – industrial standard |
|
Optical
encoder for motor position correction internal to the probe base |
|
Capability
of mapping vertical-bore magnets up to 89 mm |
|
Absolute
position reproducibility of 0.1 mm and 0.5 degree |
|
Mapping
sample volume of less than 1 mm3 |
|
Built-in
10 MHz reference with stability of 3 x10-9
per day |
|
RF modular extensions |
50, 100, 200, 400 mm |
|
Motor Shaft Length |
1.6 - 2.5 m |
|
Additional
extensions |
70
cm each |
|
Magnet
leg bracket |
Compatible
with all round and rectangular magnet supports |
|
Control Cable |
7.0 m |
|
RF
cable |
7.0
m |
|
Transit
case |
Ruggedized
field service packaging |
SPECIFICATIONS
- PROBE HEADS
|
Probe
body diameter
|
38
mm
|
|
Sample Offset Radius
|
8.5 mm
|
|
Total Length of Travel
|
100 mm
|
|
Internal probe tuning bridge
|
|
Individual
probes tuned to frequencies form 100 to 1,100 MHz
|
|
Probe tuning
range of +/- 200 MHz
|
SPECIFICATIONS
- SOFTWARE
|
MAPPER |
System
software for data acquisition, analysis, procedures. |
|
ANALYZER |
Extended
analysis, data editing and simulations. |
|
CONFIGURATOR |
System
hardware and software configuration. |
|
TESTER |
Test
software for the spectrometer, synthesizer, probe and
communications |
FMU
AnalyzerTM for WindowsR is a new Microsoft Windows based interface for the FMU-1200 magnetic field
mapping units. Launching of the new interface is expected to happen on
July 1st, 2006.
This project is being developed under the ISV partnership with
Microsoft and will support the following Microsoft software:. NET
integration, Windows Forms, C#, and Excel.
The
product is a porting to Windows of the standalone analysis package for
dataprocessing of robotically-acquired maps of magnetic fields in MRI
and NMR magnets. The Windows version will provide enhanced analytical
capabilities for this data.
SPECIFICATIONS
- SYSTEM
|
External
(host) computer |
Laptop
PC |
|
Internal
Computer |
Industrial
grade P-104 with IEEE-996 bus conversion |
|
Standard
Interface |
Serial
RS 232 - Null Modem |
|
External
power |
100-240
VAC, 50/60 Hz, 250 W |
|
Total
package weight |
90
lbs; 40 kG |
|
Total
Package dimensions |
34X24X13" |
SPECIFICATIONS
- OPTIONAL COMPONENTS
|
MXV-24 |
24
channel resistive shim controller, 16-bit resolution, allows
automatic calibration and convergence or room temperature shims |
|
ISS-10 |
Intelligent
super-conducting shim power supply, 10 channels, single 20A
circuit, to be interfaced via a IEEE-488 to enable automatic
calibration and convergence of super-conducting shims |
|
|