Applications
and Advanced Features
Animal
Imaging
Micro-Imaging
Solids
and Diffusion
Image
Gallery
Gradient
Linearity Definitions
Gradient
Design Principles
Additional
Information
|
MICRO-IMAGING
SHIELDED GRADIENTS - BFG SERIES
Most
frequently applied to vertical bore magnets used for spectroscopy, RRI's
microscopy gradients operate at strengths of 1,000 mT/m, with very high
gradient uniformity and short rise times. Microscopy gradients are fully
compatible with industry standard shim systems, including RRI's serial
and matrixshim RT shims.
Please
click here to view
images obtained with RRI's micro-imaging gradient
coils.
RRI's
gradients for microscopy can be operated with most gradient amplifiers
and are compatible with the high stability BFH gradient
amplifiers.
|
MODEL
BFG |
73/45-1000 |
73/55-500 |
|
Maximum
current input [A] |
100 |
80 |
|
Maximum
voltage input [V] |
100 |
100 |
|
Maximum
Gradient strength [mT/m] |
1,000 |
500 |
|
Duty
cycle [%] and gradient strength [mT/m] |
4% |
1,000 |
500 |
|
50% |
250 |
100 |
|
100% |
100 |
50 |
|
Rise
time, 0.5% to 99.5% [ms] |
65 |
65 |
|
DSV
[mm] |
30 |
45 |
|
Field
linearity [+/-%] |
4 |
4 |
|
Z0
compensation coil |
Yes |
Yes |
|
Shielding |
Yes |
Yes |
|
Second
order shims |
External |
External |
|
Extended
2nd, 3rd and 4th order shims |
External |
External |
|
OD
[mm] |
73 |
73 |
|
ID
[mm] |
45 |
55 |
|
Cooling |
Water |
Water |
|
Monitors |
Temperature |
PT
100 or T |
|
Water
flow |
Interlock
required |
Please
see also specifications of compatible shim
systems. Special
gradient systems for solids and diffusion studies are also
available.
|
Micro-imaging
Gradient
Model BFG-73/45-1000
45